| Raw Water : City Water < TDS 100 RO Water : AAMI Standard for Haemodialysis SPECIFICATIONS : RAW WATER PUMP CONFIGURATION : ONE RUN AND ONE STANDBY Make : Grundfos, Denmark Type : Horizontal multi-stage centrifugal pump Motor : Power Input 240V / 415V @ 50Hz Capacity : ≥ 60 LPM Material : Full S/Steel parts in fluid path Control : Electro mechanical pressure control R.O PUMP Make : Grundfos - Denmark Type : SUS and Vertical Multi Stages, Motor : Power Input 240V / 415V @ 50Hz Material : Full S/Steel parts in hydraulic fluid path PRE-TREATMENT TANK Tank Material : Fiber Glass with Vinylester lining – Max.150 psi Digital Autotrol Controller : 24 Hours 7 Days Timer for auto backwash GUARD FILTER Double Guard Filter Design Housing : 20’’ polypropylene Housing Pore Size : 1 micron cartridge filter |
BACTERIAL FILTER FOR EMERGENCY BACK UP ( Bypass Mode ) Housing : 10’’ polypropylene Housing Pore Size : 10’’ 0.2 micron bacterial filter R.O MEMBRANE OF THIN FILM COMPOSITE POLIMIDE Remove - 95% of all total dissolved inorganic matter - 99% of all organics matter greater than 300 molecular weight - 99% of collate, micro-organism and pyrogens R.O MEMBRANE OF THIN FILM COMPOSITE POLIMIDE Remove - 95% of all total dissolved inorganic matter - 99% of all organics matter greater than 300 molecular weight - 99% of collate, micro-organism and pyrogens SAFETY FEATURES 1) Built-in ELCB and MCB for all mains circuits for maximum safety. 2) Pre-low raw water tank level with audio-visual alarm. 3) Manual select Raw Water Pump ( Pump 1 - Pump 2 ) 4) Automatic system shutdown during low raw water tank level with audio-visual alarm. 5) Automatic system shutdown at low feed pressure 6) Digital conductivity meter. |